Performance Verification
Performance Verification
4Ć4
TDS 310, TDS 320, and TDS 350 Test Record
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
Performance Test Minimum Incoming Outgoing Maximum
DC Voltage Measurement Accuracy
CH1 VOLTS/DIV 1 V
200 mV
50 mV
1
50 mV
2
n at 50 mV
10 mV
5mV
+97.1 V
+8.28 V
-581 mV
-881 mV
+286 mV
+54.6 mV
-982 mV
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
+98.9 V
+8.52 V
-619 mV
-919 mV
+314 mV
+65.4 mV
-998 mV
CH2 VOLTS/DIV 1 V
200 mV
50 mV
1
50 mV
2
n at 50 mV
10 mV
5mV
+97.1 V
+8.28 V
-581 mV
-881 mV
+286 mV
+54.6 mV
-982 mV
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
__________
+98.9 V
+8.52 V
-619 mV
-919 mV
+314 mV
+65.4 mV
-998 mV
Analog Bandwidth
CH1 42.5 mV __________ __________ N/A
CH2 42.5 mV __________ __________ N/A
Long Term Sample Rate and Delay Time Accuracy
-2.0 Div __________ __________ +2.0 Div
Edge Trigger Sensitivity, DC Coupled
Main Trigger
Main Trigger - Falling
stable trigger
stable trigger
__________
__________
__________
__________
N/A
N/A
1
Generator set at -0.6 V.
2
Generator set at -0.9 V.