11
Switching the central unit on/off
After connection to the mains, the device switches on automatically and the
ON button ᕤ lights up.
̈ Press the ON button ᕤ to switch the MO 2000 CU central unit on or off.
Note:
If the power supply is interrupted for several seconds (e.g. due to a power
failure), the device switches on automatically even if it was switched off
before.
Adjusting the sensitivity
With the microphone connected and both the gain switch ᕡ and the level
control ᕢ set to 0 dB, the specified sensitivity (mV/Pa) is obtained.
̈ Use the gain switch ᕡ to increase the output voltage level by 20 dB
(tenfold increase) or 40 dB (hundredfold increase).
̈ Use the level control ᕢ to steplessly adjust the output voltage level
between –15 dB and +15 dB.
Overload indication
The OVERLOAD LED » indicates both positive and negative voltage peaks.
If the OVERLOAD LED » lights up:
̈ Use the gain control ᕡ and/or the level control ᕢ to reduce the output
voltage level so that the OVERLOAD LED » no longer lights up.