Section 4 - Operating Your Moisture Analyzer
Output Formats
Data output can be initiated in one of three ways: 1) By pressing the Print
button; 2) Using the Print Interval feature; 3) Sending a print command (P)
from a computer.
RS232 Commands
All communication is accomplished using standard ASCII format. Only the
characters shown in the following table are acknowledged by the Moisture
Analyzer. Invalid command response "ES" error indicates the Moisture
Analyzer has not recognized the command. Commands sent to the Moisture
Analyzer must be terminated with a Line Feed or carriage return-line line feed
(CRLF). Data output by the Moisture Analyzer is always terminated with a
carriage return - line feed (CRLF).
1N/C
2 <- Data Out (TXD)
3 -> Data In (RXD)
4&6 Pins 4 and 6 are connected together.
5 Ground
7 -> Clear to send (CTS)
8 <- Request To Send (RTS)
9N/C
RS232 COMMAND TABLE
V Print SR version
Esc V Print S/N (unit ID)
? Print current mode.
TIME Print Current Time.
DATE Prints Current Date.
P Print elapsed time and present reading.
Command
Character Description
4.5 RS232 Command Table
RS232 Pin out
The following table illustrates the pin-out connections on the RS232 connector.