Connection for signals with amplitudes below 10.25V RMS
Connection for signals with amplitudes to 250 V RMS
COUNT FREQUENCIES
OPERATING INSTRUCTIONS
28
COUNT FREQUENCIES
Choose this function to use the 830-KP as a frequency counter.
Available ranges are from 0.01 to 15.00 kHz, 1 to 1500 Hz and
from 1 to 1500 CPM (Counts-Per-Minute).
To measure waveforms with amplitudes between 1 V and 10.25 V
RMS use the low level inputs. Use the high voltage connection to
read waveforms with amplitudes from 10.25 to 250.0 V RMS
1) Disconnect one or both input wires from the device to be
calibrated
2) Press the FREQUENCY/PRESSURE pushbutton until any
frequency signal (KHz, Hz or CPM) appears on the display.
3) Press the DISPLAY/SOURCE/READ push-button until READ
and KHz, Hz or CPM appear on the display
4) Press the TYPE/ENG UNITS push-button to select between KHz,
Hz or CPM on the display
5) Connect the red READ lead of the calibrator to the plus (+) input
of the device and the black READ lead to the minus (-)
OPERATING INSTRUCTIONS