ResMed S9 VPAP Tx Fitness Electronics User Manual


 
18
Parameter Modes Range Description
CPAP
AutoSet
VAuto
S
ST
T
PAC
iVAPS
ASV
ASVAuto
Easy-
Breathe
(US Only)
9
On / Off
Default: On
Enables Easy-Breathe comfort
feature.
Max
Pressure
9
Min–20 cm H
2
O,
0.2 cm H
2
O
increments
Default:
20 cm H
2
O
Sets the upper limit of
treatment pressure.
Min
Pressure
9
4–Max cm H
2
O,
0.2 cm H
2
O
increments
Default:
4 cm H2O
Sets the lower limit of treatment
pressure.
Min EPAP
9
4–15 cm H
2
O
Default:
4 cm H2O
Sets the minimum EPAP
delivered by the device.
Max EPAP
9
4–15 cm H
2
O
Default:
15 cm H2O
Sets the maximum EPAP
delivered by the device.
Dependent on Min EPAP.
Options
Climate
Control
9 9 9 9 9 9 9 9 9 9
Auto / Manual
Default: Auto
Sets the type of Climate
Control.
Therapy LED
9 9 9 9 9 9 9 9 9 9
On / Off
Default: Off
Enables or disables the blue
LED.
Date
9 9 9 9 9 9 9 9 9 9
DD Mmm YYYY
00:00 (24 hr)
Sets the current date or time.
If you set a new date or time
that occurs in the past then an
‘Invalid date/time, data exists
for this period’ message is
displayed.
Time
9 9 9 9 9 9 9 9 9 9
Configuration
Language
9 9 9 9 9 9 9 9 9 9
English/ Français/
Deutsch/
Español/
Português
Default: English
Sets the display language.
Restore
factory
defaults
9 9 9 9 9 9 9 9 9 9
Yes / No Resets machine default settings
(except for language, date and
time).
Pressure
units
9 9 9 9 9 9 9 9 9 9
cm H
2
O / hPa
Default: cm H
2
O
Sets pressure unit.
Temperature
units
9 9 9 9 9 9 9 9 9 9
°F / °C
Default: °F
Sets temperature unit.
Height units
9 9 9 9 9 9 9 9 9 9
cm / inches
Default: inches
Sets height unit.