SECTION 2—OPERATION
MA90 Series 18 Part No 1148139
Mode Button (MA90Z)
1. PresstheModebutton()toselecttheMaxInflateModeortheLowAirLossmode.
•MaxInflateMode‐Inthismode,themattressinflatesrapidlytomaximum
firmness(pressurizedto35±6mmHg).Aseriesofbeepssoundeverythree
minutesasareminderthattheMaxInflatemodeisactive.After30minutes,the
MaxInflatemodedeactivates andthecontrolunitdefaultstotheprevioussetting.
NOTE:Ittakes510minutesfortheMA90Zmattresstoinflatefully(inflationtimedepends
onsizeofmattress).
NOTE:MaxInflatemodecanbemanuallydisengagedbypressingtheMaxInflatebutton.
ThiswilldeactivatetheMaxInflateLED.
NOTE:ItisrecommendedthatMaxInflatesettingbeusedduringpatientingress/egress,
patientwoundcare,patientturningorpatientcleaning.
•LowAirLossMode‐Inthismode,themattressgoesintoondemandlowairloss
reliefmode.
Max Inflate Button (MA95Z)
1. PresstheMaxInflatebutton()toselectthemaxinflatemodeandinflatethe
mattressrapidlytomaximumfirmness(pressurizedto35±6mmHg).Aseriesof
beepssoundeverythreeminutesasareminderthattheMaxInflatemodeisactive.
After30minutes,theMaxInflatemodedeactivatesandthecontrolunitdefaultstothe
previoussetting.
NOTE:Ittakes2560secondsfortheMA95Zmattresstoinflatefully(inflationtimedependson
sizeofmattress).
NOTE:ItisrecommendedthatMaxInflatesettingbeusedduringpatientingress/egress,patient
woundcare,patientturningorpatientcleaning.
2. Performoneofthefollowing:
MA90Z‐PresstheMaxInflatebuttonagaintomanuallydisengageMaxInflate
mode.
MA95Z‐SelectadifferentsettingtoexitMaxInflatemode.
NOTE:ThiswilldeactivatetheMaxInflateLED.